Computer and controller two-way communication control equipment to implement a comprehensive monitoring.
Automatic control, display single furnace or double furnace heating temperature.
Automatic measurement, showing the single furnace or double furnace coal sample shrinkage.
Automatically draws the curve of furnace temperature over time.
Automatically plot the volume curve of a double furnace or a single furnace;
Automatically fill every 10 minutes before and after the furnace temperature.
In the volume curve of the inflection point alarm, to remind the laboratory personnel on the equipment to open, stop, silence, run and other operations.
Manually check the glue surface data, through the keyboard to enter the computer to save.
Automatically draw the upper and lower gouging curves and calculate the glue surface and the maximum thickness.
Automatically calculate the test results and print out the test records and reports.
Store the previous test data for the query print.
Can be controlled before and after the furnace heating rate can also be individually controlled.
JC-8 microcomputer glial layer analyzer for the anthracite glial layer index (the maximum thickness of the glial layer Y, the final degree of shrinkage X and the volume curve type of three indicators) determination of special equipment. In line with GB / T479-2000 "anthracite glial layer index determination method" requirements. The instrument has advanced technology at home and abroad, simple operation, accurate data analysis and so on. The controller is composed of two separate systems, can control two sets of heating furnace, can also be a separate control of a heating furnace. It can be widely used in coal, metallurgy, chemical, geological exploration and other departments.
JC-8 microcomputer glial layer tester using Windows98 and above operating system, human-computer interaction, that is, school, menu and shortcut keys the perfect combination of mouse and keyboard operation, flexible and convenient. The instrument can automatically calculate the test results, accurate and reliable. Color printout test record and test report. Store the previous test data for inquiries, print. Dual interface (computer and controller) control display, easy to analyze and repair. Automatic control and description of the glial layer index in the determination of furnace temperature with time, glial layer thickness, shrinkage with the heating process of the whole process, automatically and accurately obtain X, Y value, and the volume curve type and coke technical characteristics and other indicators, And the process of the curve and the results printed out.
Power supply voltage: AC220V ± 5,50Hz
Temperature range: 0-999 ℃
Control power: ≤ 5KW
Controller accuracy: 0.5 level
Accuracy: 160 min Draw length 160 ± 2mm
Temperature control: 30min to 250 ℃
250 ° C at 3 ° C per minute
Timing error: ± 30 seconds / 24 hours
Temperature resolution: 1 ℃
Equipped with K value thermocouple
Temperature error: ± 3 ℃
Control current: up to 12A
Work: can work continuously
Record the drum: the quartz record
Line speed: 1mm / min
Measurement accuracy: in line with GB / T479P-2000 requirements