Installation method of microcomputer full automatic calorimeter
2017-08-24
Hebi Huanuo Electronic Technology Co., Ltd is the coal laboratory equipment, coal analysis instrument, computer, microcomputer automatic sulfur tester calorimeter, fluorine chlorine analyzer, bond index measuring instrument, glial layer measuring instrument, coal ash, coal quality analyzer instrument, coke laboratory equipment products such as professional production and processing the company, has a complete and scientific quality management system. Hebi Huanuo electronic technology limited company's good faith, strength and quality of products recognized by the industry. Welcome friends from all walks of life to visit, guide and business negotiation.
Hebi Huanuo electronic technology limited company production of coal testing equipment, microcomputer automatic calorimeter, automatic calorimeter in many parts, which is the main instrument of oxygen and oxygen bomb, the content of this chapter, we coal laboratory equipment and technical personnel of the oxygen instrument installation and use are introduced in this paper, hope that the new and old the customer to see the general understanding, skilled operation:
1, before the installation, carefully check whether the components are fastened, the appearance of damage and damage marks, all scattered parts to be intact.
2. Connect the oxygenator with the oxygen pressure relief valve and fasten all the nuts.
3, open the total valve of the oxygen cylinder, adjust the pressure reducing valve, and use the barometer to show 2.8-3.0MPa. At this time, the whole gas path should not leak. Otherwise, it should be checked until normal.
4, the world fully at this time should not leak the oxygen, and operate freely, with the valve on the pressure gauge consistent indicating the pressure out of the oxygenator on the table.
5, open the total valve of the oxygen cylinder (the pressure in the bottle should be greater than 5MPa), place the oxygen bomb on the positioning seat of the aerator, turn the oxygen warhead to the oxygen nozzle of the aerator, press the oxygen handle downward, and oxygenate at 2.8-3.0MPa pressure for 30 to 40s, then slowly fly and loose the handle to take out the oxygen bomb.
Oxygen meter microcomputer automatic calorimeter must be installed correctly after oxygen bomb can be normal use, the test can be completed, so oxygen instrument installation is very important, determines whether the test can be completed, whether that is accurate detection, new and old customers first to learn oxygen calorimeter instrument installation. At the same time, our company can provide quality testing instruments also provide instrument accessories, supplies, equipment maintenance, Hebi Huanuo coal equipment selection, all in one service! You buy the rest assured that with the comfortable!
Computer full automatic calorimeter similar products jump:
ZDHW-7000E high precision microcomputer full automatic calorimeter
ZDHW-8000A high precision microcomputer full automatic calorimeter
ZDHW-8000D microcomputer full automatic calorimeter (one machine and double control)
ZDHW-9000 microcomputer full automatic calorimeter
ZDHW-9000A microcomputer full automatic calorimeter
ZDHW-9000B microcomputer full automatic calorimeter
ZDHW-9000C microcomputer full automatic refrigeration calorimeter
Note: the automatic calorimeter of every microcomputer in our company will be tested by the Hebi quality and Technical Supervision Bureau and issued a quality inspection report. Please be assured of the use of the report.