Conditions of use
(1) temperature: 0 to 40 ℃
(2) relative humidity: less than 80%
(3) power supply: voltage: 220 + + 22V
Frequency ratio: 50 + / - 0.5 Hz
(4) carrier gas: helium, purity is higher than 99.8%.
(5) power: not less than 4.5KW.
(6) no strong magnetic field interference
Principle of measurement
A certain amount of coal or coke sample, in the presence of alumina as a catalyst and loose agent conditions, at 1050 ℃ access to water vapor, the sample of nitrogen and its compounds all reduced to ammonia. The resulting ammonia is distilled through a sodium hydroxide solution and titrated with a saturated boric acid solution to calculate the nitrogen content according to the consumption of the standard sulfuric acid solution.
Technical parameters
(1) measurement range: 0.01-50%
(2) sample size: about 100mg;
(3) time of single determination: about 35min;
(4) power consumption: 4000W
The technical requirements
(1) Repeatability (in line with GB / T 19227-2008) ≤0.08%
(2) temperature control accuracy: set value ± 10 ℃
(3) Temperature range:
Hydrolysis furnace: room temperature ~ 1200 ℃
Steam furnace: room temperature ~ 500 ℃
Distillation furnace: room temperature ~ 500 ℃