DN-800 semi-micro steaming nitrogen analyzer
  • Features
  • Range
  • Parameter
  • Adopt standard
  • Conditions of use

    (1) temperature: 0 to 40 ℃

    (2) relative humidity: less than 80%

    (3) power supply: voltage: 220 + + 22V

    Frequency ratio: 50 + / - 0.5 Hz

    (4) carrier gas: helium, purity is higher than 99.8%.

    (5) power: not less than 4.5KW.

    (6) no strong magnetic field interference

  • Principle of measurement

    A certain amount of coal or coke sample, in the presence of alumina as a catalyst and loose agent conditions, at 1050 ℃ access to water vapor, the sample of nitrogen and its compounds all reduced to ammonia. The resulting ammonia is distilled through a sodium hydroxide solution and titrated with a saturated boric acid solution to calculate the nitrogen content according to the consumption of the standard sulfuric acid solution.

  • Technical parameters

    (1) measurement range: 0.01-50%

    (2) sample size: about 100mg;

    (3) time of single determination: about 35min;

    (4) power consumption: 4000W

    The technical requirements

    (1) Repeatability (in line with GB / T 19227-2008) ≤0.08%

    (2) temperature control accuracy: set value ± 10 ℃

    (3) Temperature range:

    Hydrolysis furnace: room temperature ~ 1200 ℃

    Steam furnace: room temperature ~ 500 ℃

    Distillation furnace: room temperature ~ 500 ℃