1. The embedded system is composed of intelligent MCU, which has simple structure, reliable performance and strong anti-interference ability.
2. The test steps such as water discharge, water level setting, stirring, ignition, temperature collection, calibration, calculation and printing are completed automatically, which avoids human error and improves the accuracy and precision.
3. Supports two ignition methods: fuse and cotton thread
4. The chassis features a water level indicator, allowing real-time monitoring of the inner cylinder's water level.
5. The ignition voltage and ignition current can be observed at any time outside the chassis.
6. No need to weigh water and adjust water temperature, reduce the work intensity, improve the measurement accuracy.
7. It features a 7-inch color LCD screen with an integrated mounting design for enhanced stability. The device can be operated directly without requiring an external computer.
8. The instrument automatically announces the results after the experiment.
9. The instrument is controlled by a high-grade imported single-chip microcomputer system and the original imported stirring motor is used, which has the advantages of uniform speed, stable performance, reliable and strong anti-interference ability.
10. The unique calibration technique is adopted to make the instrument more tolerant to the environment, which can improve the accuracy of the test and ensure the stability of the instrument during the long time operation.
11. With the full Chinese menu interface, users only need to load the oxygen bomb to complete the entire experiment. The interface is intuitive, user-friendly, and easy to learn and use. The results are clear at a glance, and the operation is simpler and more humanized.
12. The instrument can display the ignition voltage and ignition current automatically and has the function of automatic diagnosis.
13. The instrument can measure the lower calorific value of the substance, the calorific value and weight of the additives input automatically calculated.
14. The process is automatically controlled to reduce human error. The printing cylinder, dry base high position, analysis base high position, analysis base low position, and received base heat value are measured.
High precision liquid crystal automatic calorimeter is suitable for measuring the calorific value of coal, coke, petroleum, raw material of cement, brick and other solid or liquid combustible materials in the power, coal, garbage calorific value, biomass fuel calorific value, petroleum, metallurgy, petrochemical, quality inspection, environmental protection, cement, paper, geological prospecting, scientific research institute and other industries.
1. Heat capacity: approximately 10,500 J/K
2. Operating environment: 0-65℃
3. Heat capacity precision: ≤0.1%
4. Measurement accuracy: meets the requirements of GB/T 213-2008
5. Test time: about 10 minutes
6. External water tank capacity: 75L (dual tanks)
7. Water tank capacity: approximately 2.3L
8. Ignition voltage: 24V
9. Ignition time: 3 minutes
10. Measurement error: better than national standard (<10 kcal)
11. Power consumption: ≤ 40W
12. Oxygen can capacity: 300ml
13. Water capacity: 80 litres
14. Oxygen pressure: 2.8~3.2 MPa
15. Pressure resistance: 20MPa
16. Temperature range: 0~40℃
17. Response time: <4 seconds
18. The resolution was 0.0001℃.
19. Linearity: <0.08% per 5℃ temperature rise
20. Temperature measurement error: ±0.003℃ for every 5℃ temperature rise range
21. Power supply voltage: AC220V±10%
22. Humidity: ≤80%
23. Ignition voltage: AC24V
24. Ignition time: 5S
GB/T483-2007 General Provisions for Experimental Methods of Coal Analysis
GB/T 213-2008 "Determination of calorific value of coal"
GB/T 384-1981 Determination of calorific value of petroleum products
GB/T 30727-2014 "Determination of calorific value of solid biomass fuel"
GB/T14402—2007 "Determination of combustion heat value of building materials and products"
GB/T 30991-2014 "General Technical Conditions for Intelligent Oxygen Bomb Calorimeters"
JC/T 1005-2006 "Determination of calorific value of raw cement"
JJG 672-2001 "Verification Regulations for Oxygen Bomb Calorimeters"
ASTM D5865-2013 "Determination of calorific value of coal and coke"
ASTM D5865-2007 Standard Test Method for Total Calorific Value of Coal and Coke
ISO 1928-2009 Solid Mineral Fuel-Determination of Total Value and Calculation of Net Heat Value by Bomb Calorimeter
CEN/TS 14918 "Method for the Determination of the Calorific Value of Solid Biofuels"
BS EN 15400-2011 "Solid recovered fuel-calorific value test"
IS: 1350-1970 Method for the Determination of Coal and Coke
GJB5891.29-2006 Part 29 of the Test Methods for Explosive Agents, Determination of Combustion and Explosion Heat, Isothermal Method