    CO, CO2, H2S, O2, SO2, NH3 and other test tube of the basic determination of the principle of linear colorimetric method, that is, the measured gas through the test tube and the indicator glue occurred color reaction, the formation of color layer (color column), color layer length And the measured gas concentration is proportional to.

    Storage conditions
    This product should be kept in a cool dry place, is strictly prohibited sunlight, save the temperature does not exceed 40 ℃, glass products, carefully put.

    test methods
    All kinds of test tube can be used with CZY-50 gas test tube cylinder positive pressure sampler and other supporting the use of.
    After the sampler is rinsed with air, take a certain volume of field air, cut the ends of the test tube, and connect the lower end of the test tube (one end of the concentration scale "0" with a short hose to the sampler ) On the outlet, according to the provisions of uniform time through the test tube, and then according to the test tube color column (or color ring) on the top of the number of instructions directly read the measured gas concentration.


  • Than long-type gas detection tube measurement range can be customized according to customer requirements!

    The gas detection tube is a simple detection device used to determine the concentration of a gas in an environment. The color change in the measured environment is indicated by the color change occurring by the reaction of the detection agent with the gas to be detected. With a simple operation, the use of a wide range of detection is indeed true, strong real-time and economic characteristics.

    The test range is typically in ppm. Accuracy is still good, the accuracy of 5-10%. Usually used for gas environmental safety limits, plant environmental gas limit, and individual industrial emissions, exhaust, pollution sources and other sources of pollution checks, gas leak detection, there are some industrial processes for periodic testing.


  • Name model Chemical formula Measuring range Color before and after the reaction changes Sampling volume (mL) Inflatable time (second) Term of validity(year)
    Sulfur dioxide SO2 0.0002-0.01% purple→Khaki 50 100 2
    0.02-0.1% 50 100