Sub column

Rules

Hebi Huanuo electronic computer automatic calorimeter installation
Respected users:
Hello!
Thank you for your patronage of our products.
In order to successfully carry out the equipment debugging work, ensure the equipment put into use as soon as possible, please be sure to use your organization and coordination, timely prepare the following according to the requirements provided by our company, if you have any questions please call 0392-2234888 or 2234666.
Note: if the debugging conditions do not have or can not meet the requirements, our company personnel have the right to refuse to carry out the commissioning work.
Thank You for Your Cooperation!
1. Laboratory conditions:
A, equipped with air conditioning (air outlet can not direct blowing apparatus), room temperature (5 ~ 35 DEG C) is appropriate, the relative humidity is less than 85%;
B and instruments should be placed in a place without direct sunlight or air blows. There should be no obvious air convection in the room. The indoor temperature should be relatively constant, which is in line with the national standard requirements. The temperature change in half an hour should be less than 1 degrees, and the temperature change within 1 days should be less than 5 degrees.
C, room should have solid, stable, horizontal cement, marble or other working table (recommended size: length and width of 3 x 0.8 x);
D, the best indoor water (cleaning oxygen bomb);
There should be no strong magnetic field, infrared emission source, corrosive gas and vibration source in E.
2. Power supply:
A, power supply voltage: AC220V + 10%; 50Hz
B, the power of the instrument is 180 watts.
C, power supply must be well grounded;
E, if the power supply voltage instability should be equipped with a stable voltage power supply of not less than 500VA.
3. Other items:
A, about 90 liters of distilled water, pure water or deionized water, 500ml or a glass beaker.
B, equipped with an oxygen bottle, the pressure of oxygen in the bottle should be greater than 5Mpa, oxygen purity >99.5% (no use of electrolysis oxygen, industrial oxygen electrical welding of oxygen on the line).
Calorimeter series products:
ZDHW-7000E microcomputer full automatic calorimeter
ZDHW-8000 high precision microcomputer full automatic calorimeter
ZDHW-8000D microcomputer full automatic dual control calorimeter
ZDHW-9000 microcomputer full automatic calorimeter (automatic lifting of oxygen bomb, automatic oxygen filling, automatic release of air)
ZDHW-9000A microcomputer full automatic calorimeter (automatic lifting of oxygen bomb and automatic oxygenation)
ZDHW-9000B microcomputer full automatic calorimeter (automatic lifting of oxygen bomb)
ZDHW-9000C microcomputer fully automatic refrigerant calorimeter
Note: the automatic calorimeter of every microcomputer in our company will be tested by the Hebi quality and Technical Supervision Bureau and issued a quality inspection report. Please be assured of the use of the report.
To recommend the rapid sulfur measuring instrument the nation's leading coal laboratory equipment manufacturers, professional testing equipment, coal microcomputer automatic calorimeter and microcomputer, Hebi Huanuo Electronic Technology Co., Ltd. is a professional manufacturer of domestic leading national coal laboratory equipment manufacturers professional testing equipment of coal - series of products, if you have this the demand, please contact our question, Tel: 0392-2234888